Use and maintenance of sonics ultrasonic breaker probe

1. Ultrasonic breaker probe use precautions

Do not turn on the power when the main unit is not connected to the converter.

The threaded probe must not be actuated if the tip is not replaced and the extension or tapered probe is not installed.

Do not allow the probe tip to vibrate in the air for more than 10 seconds (no load for 10 seconds).

Do not allow anything other than the sample to touch the vibrating probe.

2. Maintenance of ultrasonic breaker probe:

Proper storage of the probe is necessary for the ultrasonic breaker to operate reliably. Prolonged use of probe erosion caused by cavitation will result in lower output power and lower crushing efficiency. The smoother the probe tip, the higher the power delivered to the sample. Any erosion that occurs on the probe will further accelerate the erosion rate.

Therefore, we recommend that the probe tip be inspected once every 5 or 6 hours, if necessary with a polishing cloth or sandpaper. This maintenance can be repeated until the amplitude is set to 100, the probe is outside the sample, and the power meter reading is less than 15 watts (small probe is less than 5 watts), indicating that the probe needs to be updated.